Entry time:
Mon Mar 31 13:09:19 2025
Author
*
:
Subject
*
:
> Calibration and resolution check > > CLOVER detectors were calibrated in energy with 60Co and 137Cs > 60Co source: 2530, 8.81E4Bq (16/03/2010) > 137Cs source: CD393 (box), ? > Sources were placed at the center > > Resolution(FWHM@1333keV): > G7 Blue:2.67keV > G7 Green: 2.84keV > G7 Red: 2.77keV > G7 Black: 2.64keV > D4 Blue: 3.60keV > D4 Green: 3.35keV > D4 Red: 3.41keV > D4 Black: 4.85keV (!?) > > Efficiency calibration with 152Eu > Source place at the center > Source: 272 (09-7011), 1.062E4Bq (error:1.9%)@16/03/2010 > File: 160719_2201_152Eu_028.dlt > Online: 160719_2201_152Eu_028.root > > Attached 152Eu spectrum