AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  BRIKEN  ELOG logo
This is a draft message, edit and submit it to make it permanent  
Message ID: 61     Entry time: Thu Aug 18 22:39:35 2016
Author: NT Brewer 
Subject: New preamps arrived at Oak Ridge and tested working. 

Collaborators,

Attached is a phone photo of example signals from the new preamps with 3 3He tubes connected to a scope. Scope channels 1,2,3 (color y,c,m respectively) are preamp channels 1,2,3 while 4 (green) is open. Trigger is on channel 3, at 500mV/div and other channels are at 500,200,and 20 mV/Div for channels 2,1,4 respectively. Time scale, though glared out should be 10us per div ( although we saw longer rise times as well). 

 

Signals were all good for all channels. Coincident with real signals were other negative signals ~15x (750mV/50mV magenta/green) smaller and with a different shape. No positive signals were observed in any channel. 

 

This picture is representative of the 

 

 

 

Attachment 1: IMG_20160818_145801.jpg  4.602 MB  | Hide | Hide all
IMG_20160818_145801.jpg
ELOG V3.1.4-unknown