Both COVER detectors are installed in place. See photos. Detector D4: left of the beam Crystal distribution: BLUE-BLACK RED -GREEN Detector G7: right of the beam Crystal distribution: BLUE-BLACK RED -GREEN Cables connected. Some work on noise minimization. Resolution improved for D4-Black which was poor on July. Otherwise similar behaviour. See spectra with 137Cs and 60Co sources ta the center. File:161011_1805_CLOVER_CsCo.root We are using the calibration from July which works surprisingly well for all except G7Red (explains double peak in spectrum) More checks to be done.