Entry time:
Mon Mar 31 13:08:02 2025
Author
*
:
Subject
*
:
> Both COVER detectors are installed in place. > See photos. > > Detector D4: left of the beam > Crystal distribution: > BLUE-BLACK > RED -GREEN > > Detector G7: right of the beam > Crystal distribution: > BLUE-BLACK > RED -GREEN > > Cables connected. Some work on noise minimization. > Resolution improved for D4-Black which was poor > on July. Otherwise similar behaviour. > > See spectra with 137Cs and 60Co sources ta the center. > File:161011_1805_CLOVER_CsCo.root > We are using the calibration from July which > works surprisingly well for all except G7Red > (explains double peak in spectrum) > > More checks to be done.