Entry time:
Mon Mar 31 13:11:48 2025
Author
*
:
Subject
*
:
> Background measurement over night with CLOVER and VETO detectors > > ADC #7 added to crate 1 for veto detectors: V1A7 > VETO detectors HV off! > > Start: 23:45 > Stop: 10:56 > File: 161012_1056_bck_noise.root > > We see a very strong noise in the BRIKEN spectrum!!! > After some work we find that this was a problem > with configuration file: when editing the file > we put the calibration sheet at the beginning this > provoked a mixing of CLOVER and 3He spectra.