Subject: 252Cf 
New BRIKEN energy calibration


We check the 3He tubes energy calibration with 252Cf source
We Found that only one tube He006 was slightly shifted and correct it.

Check measurement:
File: 161014_0113_252Cf.root

See figure
Attachment 1: 252Cf-161014.pdf  38 kB  Uploaded Fri Oct 14 05:07:59 2016  | Hide | Hide all