AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
BRIKEN
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Fri Oct 14 04:42:40 2016
Author
*
:
Subject
*
:
Change of main power line for MPOD and CLOVER preamp Because the incident yesterday with the NIM crate, we try to separate the CLOVER from the rest of the equipment. So we put the MPOD, which powers the CLOVER but also the 3He tubes and the VETO detectors, and the preamp power supply in a different line. We check if this has an impact on the noise in either CLOVER or 3He tubes. We measure with the 137Cs-60Co source: File: 161014_1241_137Cs60Co.root After inspecting the spectra we conclude that the impact is not appreciable. See figures 662keV resolution: D4Black: 3.16keV D4Red: 2.49keV D4Green: 3.57keV D4Blue: 3.40keV
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
D4All-CsCo-zoom-newlineCL.pdf
Original size: 842x595
Attachment 2:
BRIKEN-newlineCL.pdf
Original size: 842x595
Attachment 3:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown