Found D2000 #1 activated and at 250C (25W)
to prevent getting dirty when other NEGs
are activated. See previous entries.
Attempted activation of D2000 #2 using T
390C , rise time 30 m, hold time 90 m
After 30 m temperature inside chamber
increased by 5C. 390 C corresponds to 75
W.
Attempted activation of D2000 #3 while #2
is holding temperature. Same parameters as
above except T 400.
After 30 min, temperature in chamber up by
10 more C to 110.
Attempted activation of #4 as above.
Set #2 then #3 then #4 to 250 C after hold
time is over to protect them. All D2000
activated. Max temperature on Kapton
bundle 110 C.
Attempted activation of 2xUHV1400 linked
to top C4, channel 2.
No thermocouple available. Used power 200
W (2x100) rise time 30 m and hold time 90
min. Upon reaching 200 W no effect on
chamber interior temperature. Increased to
300 W. Temperatures increased to 120C max
roughly. Acceptable.
Had to decrease hold time to 60 m to avoid
remaining in cave alone for too long.
Likely okay with 300 W (450 C?). Left on
150W (400C?) to protect from subsequent
activations tomorrow.
Because cables are unlabeled it's not
clear how far the activated UHV1400 are
from the detectors. If close, this recipe
works. If far, we may have to use lower
power for other modules. |