Mon 14 Sep 09:50 Jordan moved the TC08 picolog 2 away from the chamber and built a makeshit heatshield. Picolog 2 was facing a slit in the cloth thorugh which the scroll pump flexible hose is connected. Baking restarted in automatic mode from 80 C 12:09 Baking progressed OK - see attached. ~13:00 CJC 2 getting too high again. Baking went off. Jordan caught it and temporarily set in manual mode. Moved picolog 2 further away. Increased amount of heatshield. Restarted in auto 17:30 Baking continues OK - see attached. Will leave overnight. 19:30 Baking continues OK, T=124C , see attached. However rates shown on Baking.x program appear nonsensical. Too high, too low, sometimes they all go to zero. Clearly a bug in the code. Should not affect baking for now. Will try to debug.