> Mon 14 Sep 09:50 > > Jordan moved the TC08 picolog 2 away from the chamber and built a makeshit heatshield. > Picolog 2 was facing a slit in the cloth thorugh which the scroll pump flexible hose is connected. > Baking restarted in automatic mode from 80 C > > 12:09 > Baking progressed OK - see attached. > > ~13:00 > CJC 2 getting too high again. Baking went off. Jordan caught it and temporarily set in manual mode. > Moved picolog 2 further away. Increased amount of heatshield. Restarted in auto > > 17:30 > Baking continues OK - see attached. > Will leave overnight. > > 19:30 > Baking continues OK, T=124C , see attached. > However rates shown on Baking.x program appear nonsensical. Too high, too low, sometimes they all go to zero. > Clearly a bug in the code. Should not affect baking for now. > Will try to debug.