Entry time:
Tue Nov 16 12:05:13 2021
Author
*
:
Subject
*
:
<p>CARME baking with heating tape continues</p> <p>Left overnight at 100C. Turned up this morning at 9:00 to 120C.</p> <p>All good except detector strain relief seems to be going much higher than detector (no obvious reason) and then cool down. Proximity to flange / heating tape likely. One potential solution would be to push the motor in, putting some distance between detector / strain relief and outside flange.</p> <p> </p> <p>13:30 - Increased channel 1 to 130 (long tape) and channels 5 and 6 to 140 (bottom flange heaters). Left rest as was.</p> <p>15:20 - Increased channel 2 and 3 to 130. No obvious effects except Magnet pipe goes up</p> <p>16:00 - Increased channel 1, 2, 3 to 140. Increased channels 5 and 6 to 150 C.</p> <p> </p> <p>Progress can be followed on</p> <p>http://web-docs.gsi.de/~lestinsk/baking.php</p> <p>http://web-docs.gsi.de/~lestinsk/vacplot.php</p>
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
CARMEtemps.png
Original size: 1000x710
Attachment 2:
20211116_132048.jpg
Original size: 4000x2250
Attachment 3:
20211116_155138.jpg
Original size: 4000x2250
Attachment 4:
20211116_155857.jpg
Original size: 4000x2250
Attachment 5:
Drop attachments here...
Draft saved at 00:00:00