<p>CARME baking with heating tape continues</p>
<p>Left overnight at 100C. Turned up this morning at 9:00 to 120C.</p>
<p>All good except detector strain relief seems to be going much higher than detector (no obvious reason) and then cool down. Proximity to flange / heating tape likely. One potential solution would be to push the motor in, putting some distance between detector / strain relief and outside flange.</p>
<p> </p>
<p>13:30 - Increased channel 1 to 130 (long tape) and channels 5 and 6 to 140 (bottom flange heaters). Left rest as was.</p>
<p>15:20 - Increased channel 2 and 3 to 130. No obvious effects except Magnet pipe goes up</p>
<p>16:00 - Increased channel 1, 2, 3 to 140. Increased channels 5 and 6 to 150 C.</p>
<p> </p>
<p>Progress can be followed on</p>
<p>http://web-docs.gsi.de/~lestinsk/baking.php</p>
<p>http://web-docs.gsi.de/~lestinsk/vacplot.php</p> |