Message ID: 168     Entry time: Tue Jan 11 12:45:15 2022
Author: TD, JM, OH 
Subject: Coolant manifold installation + detector pin tests 
2x cooling water manifolds with 2x 8 ports have been installed - see attachments 1 & 2

'drain' ports at top of manifold to permit release of trapped air volumes. Local technicians will install a standard blank in one manifold and a small nipple valve on the 
other so that we can bleed air in a controlled way.

#Multimeter detector pin tests. 
Pins on feedthrough flanges investigated using multimeter for the top and bottom detector due to no response from bottom detector when bias was applied. Bias pins on 
correct D connectors are number 78. 

#Top detector
>Observe -18.5 kohm; between nn bias pin and pn bias pin  (multimeter 200kohm setting)
>observe open line between nn bias pin and pins 'near' pn bias pin (multimeter 200kohm setting)

>Between random pins on same D connector (nn bias pin connector) ~0.02Mohm (multimeter 2Mohm setting)
>Between random pins and unconnected on same D connector (nn bias pin connector) - open line (multimeter 2Mohm setting)
>Between random pins and bias pin on same D connector (nn bias pin connector) ~0.01-0.02 Mohm (multimeter 2Mohm setting)

>Between nn bias pin and random pin on another D connector ~-0.4Mohm (multimeter 2Mohm setting)

#Bottom detector 
>Observe open line between nn bias pin and pn bias pin (multimeter 200kohm+2Mohm; setting)
>Observe open line between nn bias pin and pins 'near' bias bias pin (multimeter 200kohm+2Mohm; setting)

>Between random pins on same D connector (nn bias pin connector) ~0.02Mohm (multimeter 2Mohm setting)
>Between random pins and unconnected on same D connector (nn bias pin connector) - open line (multimeter 2Mohm setting)
>Between random pins and bias pin on same D connector (nn bias pin connector) ~0.01-0.02 Mohm (multimeter 2Mohm setting)

>Between random pins on same D connector (pn bias pin connector) ~0.02Mohm (multimeter 2Mohm setting)
>Between random pins and unconnected on same D connector (pn bias pin connector) - open line (multimeter 2Mohm setting)
>Between random pins and bias pin on same D connector (pn bias pin connector) ~open line (multimeter 2Mohm setting)

>Between nn bias pin and random pin on another D connector ~-0.4Mohm (multimeter 2Mohm setting)

We see a similar response for both detectors except that the pn bias pin gives no response. 
We have looked to find the bias pin on nearby pins to pin 78 and on the other side of the D connector (in case of cabling error-mirror or swapped cables) but cannot find. 

4 FEE holders constructed and ready to install