Detector is not biasing as it has in previous tests. Last bias 15/1/22 see previous elogs, max leakage current ~10 uA at 150V
Detector bias increases well above 10uA for voltages as low as 30-40V. Optical windows covered, Y09 valves closed, large ion pump OFF and both ion gauges OFF produces similar results but with a slightly lower leakage current.
Grounding and pulser cables removed from FEEs and jumper attached to nn bias adaptor card - no change. Bias module swapped out - no effect. NIM rack shows expected voltages (6,12,24) on multimeter.
AIDA03 - disconnected. Behaves as expected. Get waveforms and pulser peak. Pulser peak FWHM ~ 70 channels, noise transients observed.
AIDA02 - Not producing ADC data
AIDA01,04,05,06,07,08 - Very high rates - noisy
|