08:52 Overnight the DAQ was left running to check if there was any individual radioactivity in the Si.
This is stored in R7_75
It was observed that even with no beam the detectors continued to increase in leakage current overnight but slowly - Attachment 1
We then stopped the DAQ and gave the detector a single shake.
This caused the leakage currents to decrease but not to the same level that they did last night - Attachment 2
13:45 Bias tripped as people removed the cover on the windows on the interaction chamber
Iset was still at 11uA for safety
Will leave off until they are finished
|