<p>> Initial grounding see final grounding from elog 187. ASIC settings for all - LEC slow comparator 0x64. Note on 11/02/22 1.8*L widths with same settings were aida1-4: 300,240,250,260</p>
<p>ADC data items <10k</p>
<p>Good events fluctuate 10k-20k</p>
<p>1.8*L pulser widths aida1-4: 433,400,375,375(double peak)</p>
<p>Attachments 1-4</p>
<p>> New grounding implemented. 3mm grounding cable connect fee body to adaptor board. Each adaptor board is then connected directly to the cryring ground. Attachements 10-11</p>
<p>ADC data items <10k</p>
<p>Good events fluctuate 10k-20k</p>
<p>1.8*L pulser widths aida1-4: 340,288,288,295 (reduced double peak)</p>
<p>Noise in the system also seems to have decreaed in zoomed out 1.8*L plots (not attached). Aida1 still shows large noise peak channel 32000 on 1.8*L. 1.8*W no significant change, perhaps slightly reduced.</p>
<p>Attachements 5-9</p>
<p>Screencaps for following movement tests not taken as 1.8*L and 1.8*W spectra appeared the very similar as 5-9 except for slightly different widths. </p>
<p>> Detectors moved in and out (pneumatic). Leakage current dropped by ~0.2 uA, torque now reads +2.8 previously was ~0. </p>
<p>no change to ADC data items or good events. During movement aida3 saw momentary increase in good events to ~40k before returning to normal range. Rest of fees good events stats unaffected by movement.</p>
<p>1.8*L pulser widths aida1-4: 413,313,295,306</p>
<p>> servo motor (only) moved, torque reduced back to ~0</p>
<p>no change in good events stats for any fees during servo movement. </p>
<p>1.8*L pulser widths aida1-4: 425,335,316,288</p>
<p>> Detector moved to IN position (pnuematic and servo). 19.3mm to 81.8mm on potentiometer. </p>
<p>No change to good events. During movements aida3 saw momentary increase in good events to ~41k, all other fees unaffected. </p>
<p>1.8*L pulser widths aida1-4: 302,221,305,293 </p>
<p>Noise peak at channel ~32000 on aida 1 (1.8*L) significantly reduced. little noise on other fees </p>
<p>>Detector moved to fully OUT position (pneumatic and servo). </p>
<p>no change in good events, during movement aida3 saw momentary increase to ~35k</p>
<p>1.8*L pulser widths aida1-4: 450,350,316,290</p>
<p>Noise peak at chn 32000 on aida 1 has increased to previous size. little noise on other fees. </p>
<p>> Bias increased to 120V. Leakage current stable however slightly higher ~0.1 uA</p>
<p>no change in good events. </p>
<p>1.8*L pulser widths aida1-4: 447,345,189,301</p>
<p>> Bias increased to 140V, increase in leakage current ~0.2 uA, appears stable. Increase to 150V, fluctuate in leakage current of +/-0.5uA then huge increase to >20uA trip. </p>
<p>> Fee power cycle. Bias 100V, leakage current 9.9 uA ~0.1-0.2 uA higher than previous but stable. </p>
<p>no change to good events</p>
<p>1.8*L pulser widths aida1-4: 468, 350,311, 296</p>
<p>> Some decrease seen with new grounding, however main source of noise still apparent and any decrease seems insignificant compared to changes due to position movement. </p>
<p>> Water pipes have been zipped tied so they no loger touch motors, no change in pulser widths or noise levels however. </p> |