Connected Leybold Ionivac IE514 to Leybold Combivac CM52 Vacuum Gauge Controller Rear panel - power ON POST checks OK Change parameters by select PARA select <-> to display Sen select PARA to display S-on select <-> to display HAnd select PARA and hold for 7s select CH to select CH3 select -> hold for 2s With CARME gate valve closed for c. 24h CARME chamber pressure is 6.43e-5mbar @ 13.34, 6.52e-5mbar @ 14.15 15.45 CARME Leybold turbo restarted 16.06 1.58e-6mbar, 3.5e-5mbar (RGA station), leak rate 3.6e-10 mbar.l/s 16.27 9.44e-7mbar, 3.3e-5mbar (RGA station), leak rate 3.6e-10 mbar.l/s 16.39 7.67e-7mbar 17.00 5.85e-7mbar, 2.5e-5mbar (RGA station), leak rate 2.60e-10 mbar.l/s 17.20 4.83-7mbar, 2.3e-5mbar (RGA station), leak rate 2.17e-10 mbar.l/s Baking octagon moved into position along with the bottom insulation and heaters