AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
CARME
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Wed Sep 23 13:34:32 2020
Author
*
:
Subject
*
:
Connected Leybold Ionivac IE514 to Leybold Combivac CM52 Vacuum Gauge Controller Rear panel - power ON POST checks OK Change parameters by select PARA select <-> to display Sen select PARA to display S-on select <-> to display HAnd select PARA and hold for 7s select CH to select CH3 select -> hold for 2s With CARME gate valve closed for c. 24h CARME chamber pressure is 6.43e-5mbar @ 13.34, 6.52e-5mbar @ 14.15 15.45 CARME Leybold turbo restarted 16.06 1.58e-6mbar, 3.5e-5mbar (RGA station), leak rate 3.6e-10 mbar.l/s 16.27 9.44e-7mbar, 3.3e-5mbar (RGA station), leak rate 3.6e-10 mbar.l/s 16.39 7.67e-7mbar 17.00 5.85e-7mbar, 2.5e-5mbar (RGA station), leak rate 2.60e-10 mbar.l/s 17.20 4.83-7mbar, 2.3e-5mbar (RGA station), leak rate 2.17e-10 mbar.l/s Baking octagon moved into position along with the bottom insulation and heaters
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
20200923_170441.jpg
Original size: 4128x3096
Attachment 2:
20200923_170455.jpg
Original size: 4128x3096
Attachment 3:
20200923_170415.jpg
Original size: 4128x3096
Attachment 4:
20200923_161415.jpg
Original size: 4000x3000
Attachment 5:
20200923_161402.jpg
Original size: 4000x3000
Attachment 6:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown