> > Connected Leybold Ionivac IE514 to Leybold Combivac CM52 Vacuum Gauge Controller > > Rear panel - power ON > > POST checks OK > > Change parameters by > > select PARA > > select <-> to display Sen > > select PARA to display S-on > > select <-> to display HAnd > > select PARA and hold for 7s > > select CH to select CH3 > > select -> hold for 2s > > With CARME gate valve closed for c. 24h CARME chamber pressure is 6.43e-5mbar @ 13.34, 6.52e-5mbar @ 14.15 > > > 15.45 CARME Leybold turbo restarted > > 16.06 1.58e-6mbar, 3.5e-5mbar (RGA station), leak rate 3.6e-10 mbar.l/s > > 16.27 9.44e-7mbar, 3.3e-5mbar (RGA station), leak rate 3.6e-10 mbar.l/s > > 16.39 7.67e-7mbar > > 17.00 5.85e-7mbar, 2.5e-5mbar (RGA station), leak rate 2.60e-10 mbar.l/s > > 17.20 4.83-7mbar, 2.3e-5mbar (RGA station), leak rate 2.17e-10 mbar.l/s > > Baking octagon moved into position along with the bottom insulation and heaters >