AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CARME  ELOG logo
Message ID: 212     Entry time: Sat Feb 19 14:51:58 2022
Author: JM, CB 
Subject: Burst high energy noise 

Noticed bursts of high energy noise on aida01 and aida02 during run 47 and 48.

Good events rate jumps from 10-20k range to 250k. ADC data items rate jumps from 0-1 to 2k. Data rate being written increases to ~2500 kb/s from ~20 kb/s

Effect appears short lived ~30s at a time. On 18/2/22 the asic was check loaded which appeared to remove the increased rate but due to the time taken to asic check the burst had already finished.

Asic 2 on aida01 and asic 1 on aida02 are affected. See attached plots.

Fast comparator threshold HEC currently set to 0x2 for all

Zoomed in region of aida 1 asic 2 and the same region for asic 1 to see the difference are attached.

 

Attachment 1: 2022-02-19_15-50-50-aida1-4-stat-logscale.png  167 kB  | Hide | Hide all
2022-02-19_15-50-50-aida1-4-stat-logscale.png
Attachment 2: 2022-02-19_15-47-13-aida1-2*H.png  197 kB  | Hide | Hide all
Cannot create thumbnail, please check ImageMagick installation
Attachment 3: 2022-02-19_15-49-48-aida2-1*H.png  207 kB  | Hide | Hide all
Cannot create thumbnail, please check ImageMagick installation
Attachment 4: 2022-02-20_14-01-40-aida1-2*H.png  247 kB  Uploaded Sun Feb 20 13:05:51 2022  | Hide | Hide all
Cannot create thumbnail, please check ImageMagick installation
Attachment 5: 2022-02-20_14-02-55-aida1-1*H.png  211 kB  Uploaded Sun Feb 20 13:06:06 2022  | Hide | Hide all
Cannot create thumbnail, please check ImageMagick installation
ELOG V3.1.4-unknown