<p>04:24</p>
<p>Temperature setpoint 110 K</p>
<p>Rates appeared stable throughout the hour ~10^2 peak height. Last 5 minutes rutherford peak height appears reduced, cycling from 10^1 to 10^2. If this continues with sustained low density, temperature will be cycled to 125 K and back down.</p>
<p>Merger OK</p>
<p>Data rate = 1300 kb/sec</p>
<p>Ring avg pressure = 3.1E-10 mbar</p>
<p>Temperatures OK</p>
<p>System checks OK - wr fail attached</p>
<p>Leakage current ~ 3.228 uA</p>
<p>04:33</p>
<p>Rutherford rate appears to have increased and stabilised again around 10^2 peak height.</p> |