<p>23:35</p>
<p>Temperature setpoint = 116 K</p>
<p>Detector position IN, potentiometer reads 34.89 mm</p>
<p>Rutherford rate fluctuates however is consistently aboe 100 counts peak height</p>
<p>Merger OK</p>
<p>Data rate = 1600 kb/sec</p>
<p>Leakage current = 3.23 uA</p>
<p>Temperatures OK</p>
<p>System checks OK - wr attached</p>
<p> Base Current Difference<br />
aida01 fault 0xab60 : 0xab8c : 44 <br />
aida02 fault 0x88d3 : 0x88fe : 43 <br />
aida03 fault 0x23aa : 0x23d6 : 44 <br />
aida04 fault 0x15d7 : 0x1603 : 44 <br />
aida05 fault 0x2f03 : 0x2f0f : 12 <br />
White Rabbit error counter test result: Passed 0, Failed 5</p>
<p>Understand the status reports as follows:-<br />
Status bit 3 : White Rabbit decoder detected an error in the received data<br />
Status bit 2 : Firmware registered WR error, no reload of Timestamp<br />
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR</p>
<p> </p> |