<p>06:20</p>
<p>Servo motors moved to fully out position as some work on the beam was neccessary. Beam injection had dropped from ~9e7 to around 1e7. Potentiometer reads 19.10 mm. </p>
<p>Still trying to improve target density. Current setpoint 100 K.</p>
<p>Merger OK</p>
<p>Data rate = 1600 kb/sec</p>
<p>Leakage current = 3.21 uA</p>
<p>System checks OK - wr fails</p>
<p> Base Current Difference<br />
aida01 fault 0xab60 : 0xab8e : 46 <br />
aida02 fault 0x88d3 : 0x8900 : 45 <br />
aida03 fault 0x23aa : 0x23d8 : 46 <br />
aida04 fault 0x15d7 : 0x1605 : 46 <br />
aida05 fault 0x2f03 : 0x2f0f : 12 <br />
White Rabbit error counter test result: Passed 0, Failed 5</p>
<p>Understand the status reports as follows:-<br />
Status bit 3 : White Rabbit decoder detected an error in the received data<br />
Status bit 2 : Firmware registered WR error, no reload of Timestamp<br />
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR</p>
<p>Avg ring pressure = 4.3E-10 mbar.</p>
<p>06:40</p>
<p>Target density still very low. Setpoint back to 116 K. Detectors still fully out, very low rutherford rate.</p>
<p>Attempted to move nozzle to increase density, already on maximum point. </p>
<p>06:50</p>
<p>Setpoint to 95 k. Small increase in target density, however this was temporary</p>
<p> </p> |