<p>detector motors are still out</p>
<p>'Check Clock Status', 'Check ADC Calibration' : Clock status test result: Passed 5, Failed 0</p>
<p>'Check the White rabbit decoder status' : <br />
Base Current Difference<br />
aida01 fault 0xab60 : 0xab8f : 47 <br />
aida02 fault 0x88d3 : 0x8901 : 46 <br />
aida03 fault 0x23aa : 0x23d9 : 47 <br />
aida04 fault 0x15d7 : 0x1606 : 47 <br />
aida05 fault 0x2f03 : 0x2f0f : 12 <br />
White Rabbit error counter test result: Passed 0, Failed 5</p>
<p>Understand the status reports as follows:-<br />
Status bit 3 : White Rabbit decoder detected an error in the received data<br />
Status bit 2 : Firmware registered WR error, no reload of Timestamp<br />
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR</p> |