<p>22:16</p> <p>Temperature setpoint = 113k</p> <p>Not observed noise burst again</p> <p>Target density OK, S1 between 2E-8 and 3E-8mbar, E4 between 1E-9 and 1E-8 mbar.</p> <p>Detector moving well, potentionmer between 23.1 mm and 67.43 mm, move in/move out counters the same. Correlation scalar on aida1</p> <p>Ruthrford peak varies between 10^3 and 10^2 peak height depending on detector position, noise rate also varies with position.</p> <p>Beam intensity ~1E8</p> <p>Merger OK</p> <p>Data rate ~ 1200 kb/sec</p> <p>Temperatures OK</p> <p>System checks OK - wr fails</p> <p> Base Current Difference<br /> aida01 fault 0xab60 : 0xaba4 : 68 <br /> aida02 fault 0x88d3 : 0x8915 : 66 <br /> aida03 fault 0x23aa : 0x23ee : 68 <br /> aida04 fault 0x15d7 : 0x161b : 68 <br /> aida05 fault 0x2f03 : 0x2f20 : 29 <br /> White Rabbit error counter test result: Passed 0, Failed 5</p> <p>Understand the status reports as follows:-<br /> Status bit 3 : White Rabbit decoder detected an error in the received data<br /> Status bit 2 : Firmware registered WR error, no reload of Timestamp<br /> Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR</p> <p>Ring pressure oscillation is 30 minute cycle on all areas of the ring.</p>