Tue Aug 9 13:44:16 2022
Author
*
:
Subject
*
:
14.00 CARME vacuum c. 1.8e-10 mbar - see attachment 1 open water inlet/outlet taps NIM bin (MACBs, Ortec 480 Pulser, CAEN N1419ET) switched ON USB-controlled ac mains relay switched ON - water & interlock LEDs active Require VETAR2 (previous returned to Nik Kurz) and network switch (currently with AIDA@DESPEC) Turbo pump ON - normal operation Ion pumps ON - normal operation 16.30 ML informs us that ambient temperature of CRYRING cave has been increased from c. 16 deg C to c. 21 deg C and that this is likely the cause of the change in vacuum pressure in YR09 and the rest of the storage ring. obtained VETAR2 and crate anydesk restarted
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
Capture_1.PNG
Original size: 1913x1080
Attachment 2:
20220809_163450.jpg
Original size: 4000x3000
Attachment 3:
20220809_135840.jpg
Original size: 2250x3000
Attachment 4:
20220809_135833.jpg
Original size: 2250x3000
Attachment 5:
20220809_135801.jpg
Original size: 4000x3000
Attachment 6:
Drop attachments here...
Draft saved at 00:00:00