Moved pneumatic motor in and out several revolutions
Biased detector up to 100 V, I ~ 12.4 uA ambient temp c. 21 deg C
Moving detector in increased leakge current to ~14.5 uA, moving detector out dropped leakage current back to ~12.4 uA
aida05 - pulser width 21 chn (slightly lower than yesterday)
aida01 - pulser width 124 chn
aida02 - double peak (is not currently grounded)
aida03 - 183 chn
aida04 - 212 chn
Attachment 1 1.8.L spectra p+n FEE64s
Attachment 2 per FEE64 rate spectra
Attachment 3 aida05 1.8.L, 1.8.W and rate spectra
Attachment 4 FEE64 temperatures - OK
Attachment 5 per FEE64 ADC data event stats - rates for aida01-aida04 significabtly lower than yesterday presumably due to detector bias -100V
Attachments 6 & 7 per FEE64 1.8.W spectra
Attachment 8 1.8.L spectra for n+n FEE64s
regrounded aida02 back to cryring ground - required removing star configuration of all fee grounds and making a new
aida05 - pulser width 30 chn
aida01 - pulser width 300 chn
aida02 - 220 -> double peak gone
aida03- 190
aida04 - 200
Perhaps star configuration isn't as tight as before - re tighten star section
Re tightened section -> no impact on pulser widths
Moved detector in using pneumatic -> no impact on pulser widths
Added a grounding cable from the 19 inch rack to the same CRYRING ground that the fees are connected too. Stats around 2-3k, slow comparator = 0x64, bias=100v, I=12.18 uA
aida01 ~ 260
aida02 ~ 177
aida03 ~ 178
aida04 ~ 187
aida05 ~28
Removed CRYRING ground. Fees and rack connected together still.
no change
Bias supply module is floating
ion pumps off, gauge off -> no change in waveforms, leakage current dropped to 8.7 uA
aida01 ~ 255
aida02 ~ 176
aida05 ~ 26
Removed all vacuum systems and motors power from 19 inch rack power. Connected to the rack are now: NIM bin, network switch, 2 inputs to mains relay, raspberrypi, pi dlink usb connector, VME crack
aida01 ~ 220
aida02 ~ 154
aida05 ~ 27
aida03 ~ 181
aida04 ~ 193
Moved power cable for 19 inch rack to CRYRING 'clean' power source
aida03 ~ 211
aida04 ~ 212
aida01 ~ 317
aida02 ~ 215
aida05 ~ 29
Reconnected CRYRING ground
aida01 ~ 301
aida02 ~ 203
aida05 ~ 28
aida03 ~ 181
aida04 ~ 191
Current status 17:00:
All CARME vacuum equipment is off (turbo, SIP power modules, NEG power modules, gauges) - roughing pump is on but is valved off from the turbo which is valved off from chamber
Motors and motors pi are off. Detectors are fully out and the pin has been reinserted.
Bias ON 100V, 8.5 uA. Pulser ON, all equipment connected to 19 inch rack power ON. Jumper is on link 1 for nn bias adaptor card only. |