AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
CARME
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Fri Apr 4 17:18:26 2025
Author
*
:
Subject
*
:
<p>9:40</p> <p>DSSD bias and leakage current values:</p> <p>-99.99 V</p> <p>-13.0960 uA</p> <p>Pulsar peak widths (ch FWHM)</p> <p>aida01 ~ 97</p> <p>aida02 ~ 103</p> <p>aida03 ~ 134</p> <p>aida04 ~ 131</p> <p>aida05 ~ 22</p> <p>Pretty similar to what was found yesterday</p> <p> </p> <p>15.30</p> <p>Per Elog 375</p> <p>Measured FEE64 voltages are as follows</p> <p>@ISOL 4 AIDA PSU @FEE64 aida05 nominal</p> <p>(top-botttom) (right-left)</p> <p>+6.05V +5.26V +5V</p> <p>+6.05V +5.26V +5V</p> <p>+6.05V +5.26V +5V</p> <p>-6.49V -6.26V -6V</p> <p>+7.78V +7.46V +7V</p> <p> </p> <p>Adjusted PowerStax PSU voltage pots of AIDA PSU ISOL 4</p> <p>@ISOL 4 AIDA PSU @FEE64 aida05 nominal</p> <p>(top-botttom) (right-left)</p> <p>+5.79V +4.99V +5V</p> <p>+5.79V +4.99V +5V</p> <p>+5.79V +4.99V +5V</p> <p>-6.23V -6.00V -6V</p> <p>+7.32V +7.00V +7V</p> <p> </p> <p>Attatchment 1-12 = BEFORE change</p> <p>Attatchment 13+ = AFTER change</p> <p>Pulsar peak widths (ch FWHM) AFTER CHANGE</p> <p>aida01 ~ 100</p> <p>aida02 ~ 96</p> <p>aida03 ~ 130</p> <p>aida04 ~ 130</p> <p>aida05 ~ 22</p>
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown