AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CARME  ELOG logo
Message ID: 436     Entry time: Wed Aug 30 09:33:13 2023
Author: CB, OG 
Subject: NEG cables repaired  

Repaired Neg wires for D2000 with Oemer's help.

Cable #2 is fine. 

In Cable #4, wires 1 and 2 are swapped but it's impossible this happened when the connector came apart. The pins are crimped and secured with a screw. Most likely it was sold like this and there's no difference between wire 1 and 2 (same voltage).

Edit: SAES confirmed 1&2 are interchangeable.

Attachment 1: 20230830_102519.jpg  1.606 MB  Uploaded Wed Aug 30 10:42:09 2023  | Hide | Hide all
20230830_102519.jpg
Attachment 2: 20230830_102514.jpg  1.598 MB  Uploaded Wed Aug 30 10:42:45 2023  | Hide | Hide all
20230830_102514.jpg