Entry time:
Sun Oct 15 09:35:03 2023
Author
*
:
Subject
*
:
<p>10:40 - Status as attached. Heaters setpoint/read at 250 C, 150 C, 225 C. Power stable at 53%, 41%, 63%</p> <p>10:45 - Informed HKR ramp down / pump activation is starting for CARME only. Setpoints to 200, 125, 200 C. Monitoring slow ramp down.</p> <p>14:10 - Heater temperatures reached the set points. Power stabilising. CARME internals temperature dropped to around 100 C and still falling. Waiting until later today to further reduce heater setpoint. Will probably aim for 80 C before beginning activation.</p> <p>15:45 - Heater setpoints to 175, 100, 175 C</p> <p>15:50 - Interaction chamber jacket setpoints to 100C</p> <p>18:00 - CARME temperature decreasing, not stable yet. Will check later tonight.</p> <p>20:43 - CARME internals around 85C, not quite stable still. Waiting overnight.</p>