Entry time:
Fri Apr 4 17:18:24 2025
Author
*
:
Subject
*
:
<p>Mounted water manifold on CARME with help from GSI plumber. Connected all the hoses with Swagelok self-locking connectors. Tested for water leaks - it leaks from nearly every port. The issue appear to be the connections between the manifold and the plugs, as opposed to the plugs and the hoses which do not leak.</p> <p>Will try to find a solution tomorrow.</p> <p> </p> <p>FEEs connected from bottom-left, top-left and top-right detector.</p> <p>Top right detector showed same behaviour described in previous ELOG entry. Removed all jumpers from PCBs. Beahaviour is gone.</p> <p>Removed all jumpers from flexi-rigid PCBs for bottom-left detector. Behaviour is improved, but not gone - still 13 uA at 150V. Around 3.5 uA at 120V with no conditioning - that may be the working point.</p> <p> </p> <p>Missing</p> <p>- Power cables and ethernet cables for half the fees</p> <p>- Fix leaks on water manifold</p> <p>- Fix / change / re-arrange fee holder for bottom-right detector to avoid it fouling on CARME frame.</p>