Entry time:
Tue Aug 31 11:55:11 2021
Author
*
:
Subject
*
:
Found strut bolts in the bag with spherical washers. Mounted the struts and the washers. See attach 1. Vented CARME in preparation for lift. Turbo off Pressure starts increasing without introducing leaks. Up to 10-4 mbar. IE 514 off. Removed plastic cap from leak valve. Pressure rate increases. Leaky valve? Turbo motor speed reaches 0 Hz in 1h Opened leak valve. Scroll off. Pressure to 10 mbar. Not increasing anymore. Turbo leak aperture likely to small to vent CARME volume. Disconnected scroll. Carefully let air in by tweaking flexible hose connection. Not recommended again, but should be safe given turbo fully stop and chamber nearly vented. Reached atmosphere in 10 minutes. Now waiting for lifting support to mount feet.