> Found strut bolts in the bag with > spherical washers. Mounted the struts and > the washers. See attach 1. > > Vented CARME in preparation for lift. > > Turbo off > Pressure starts increasing without > introducing leaks. Up to 10-4 mbar. IE 514 > off. > > Removed plastic cap from leak valve. > Pressure rate increases. Leaky valve? > > Turbo motor speed reaches 0 Hz in 1h > > Opened leak valve. Scroll off. Pressure to > 10 mbar. Not increasing anymore. Turbo > leak aperture likely to small to vent > CARME volume. > > Disconnected scroll. Carefully let air in > by tweaking flexible hose connection. Not > recommended again, but should be safe > given turbo fully stop and chamber nearly > vented. > > Reached atmosphere in 10 minutes. > > Now waiting for lifting support to mount > feet.