CARME |
 |
|
Message ID: 481
Entry time: Tue Jan 30 09:14:50 2024
|
Author: |
JM, TD |
Subject: |
Tuesday 30 January |
|
|
09.00 CARME IE514 5.1e-11mbar
09.15 IE514 not working - one of cables disconnected?
DSSSD #1 bias cable changed from SHV - Lemo 00.250 to SHV - 2x Lemo 00.250
DSSSD #1 & #2 LK1 removed from n+n Ohmic side bias FEE64s
aida11 HDMI cabling re-seated at MACB
10.26 DSSSD bias OK - attachment 1
Ambient temperature +16.2 deg C
10.30 ADC data item stats OK - attachment 2
aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser
10.34 per FEE64 Rate spectra - attachment 3
OK except for known issues aida05, aida07 and aida16 - all others OK
per p+n FEE64s 1.8.L spectra - attachment 4
pulser peak width 52 ch FWHM aida11- double peaking DSSSD #1 & #2
per n+n FEE64s 1.8.L spectra - attachment 7
pulser peak width 77 ch FWHM aida04 - more uniform noise across all DSSSDs
10.42 per FEE64 1.8.W spectra - 20us FSR - attachments 5-6
11:00 LK2 removed from aida01-04. Configuration for all adaptor cards/FEE modules is now the same. No other change to the setup.
12:00 FEEs power on, bias to 150 V. No change in bias observed.
12.02 ADC data item stats OK
aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser
12.07 per FEE64 Rate spectra
OK except for known issues aida05, aida07 and aida16 - all others OK
12:09 per p+n FEE64s 1.8.L spectra
We now observe a large peak around chn 30,000 on all FEEs. Perhaps an effect of some change in the ring?
pulser peak width 53 ch FWHM aida11- still observe double peaking DSSSD #1 & #2
per n+n FEE64s 1.8.L spectra
Lots of events at ~chn 65,000 - double peaking now observed on aida03 and aida04
pulser peak width 93 ch FWHM aida14
12.20 per FEE64 1.8.W spectra - 20us FSR
Perhaps bias was off while acquiring and spectra was not cleared properly. Re-do tests
13:40 ADC data item stats OK - attachment 7
aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser
per FEE64 Rate spectra - attachment 8
OK except for known issues aida05, aida07 and aida16 - all others OK
per p+n FEE64s 1.8.L spectra - attachment 9
pulser peak width 55 ch FWHM aida11- double peaking DSSSD #1 & #2
per n+n FEE64s 1.8.L spectra - attachment 10
pulser peak width 65 ch FWHM aida14 - more uniform noise across all DSSSDs
per FEE64 1.8.W spectra - 20us FSR - attachments 11-12
With help from NH updated firmware on all MACB modules to version 4/20 - see https://elog.ph.ed.ac.uk/CARME/482
Timestamps for all FEE modules now correct - attachment 13
MAC address of aida07 was found to be 41-b4-16 not 41-d4-16 as was labelled on the FEE module
re-labelled FEE64
dhcpd.conf and MIDAS startup.tcl updated, dhcpd and MIDAS 8015 server restarted - aida07 is now working OK
CAEN N1419ET internal jumper fitted - outputs are no longer floating
15:20 ADC data item stats OK
aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser - attachment 14
per FEE64 Rate spectra
OK except for known issues aida05, and aida16 - all others OK - attachment 15
per p+n FEE64s 1.8.L spectra -
pulser peak width 50 ch FWHM aida11- double peaking DSSSD #1 & #2 - attachment 16
per n+n FEE64s 1.8.L spectra
pulser peak width 67 ch FWHM aida14 - more uniform noise across all DSSSDs
per FEE64 1.8.W spectra - 20us FSR - aida14 missing from waveforms - attachments 17,18
Swapped the test inputs so that +ve is now from the inverter and -ve direct from pulser. - attachments 19,20. Observed reduced noise for when plugged directly into the pulser than when connected from the inverter.
Lowered threshold from 0x64 (1 MeV) to 0x20 (320 keV) - attachment 21, 22 - Observed increase in rates, in particular for the DSSD #1,2 pn side. Conclusion -> increased pulser width for DSSD #1,2 is real and not from the pulser
aida01 lemo to water pipe (chamber ground) connection installed
ADC data item stats OK
aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser
per FEE64 Rate spectra
OK except for known issues aida16 - all others OK
per p+n FEE64s 1.8.L spectra -
pulser peak width 48 ch FWHM aida11- double peaking DSSSD #1 & #2 - attachment 23
per n+n FEE64s 1.8.L spectra
pulser peak width 67 ch FWHM aida14 - attachment 24
per FEE64 1.8.W spectra - 20us FSR - slightly reduced 100 kHz on pn waveforms - attachments 25,26
Disconnected aida06 and replaced with the spare FEE module (mac address: 42:0d:16). Power cycled and updated config files.
ADC data item stats OK
aida15 > 100k, all others including aida06 c. 3k consistent with 50Hz pulser - attachment 27
per FEE64 Rate spectra
OK except for known issues aida16 - all others OK - attachment 28
per p+n FEE64s 1.8.L spectra -
pulser peak width 48 ch FWHM aida11- no longer observe double peaking DSSSD #1 & #2 but the pulser peaks are still broader compared to DSSD #3,4 - attachment 29
per n+n FEE64s 1.8.L spectra
pulser peak width 55 ch FWHM aida14 - attachment 30
per FEE64 1.8.W spectra - 20us FSR - greatly reduced 100 kHz on pn waveforms, dodgy FEE previously aida06 must@ve had some impact on DSSD#1,2 - attachments 31,32
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|