AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CARME  ELOG logo
Message ID: 494     Entry time: Mon Feb 5 20:15:31 2024
Author: CB, TD 
Subject: Detector move in/out signals 
17.43 setup signals from detector movement controller.

See attachment 1
                                                         NIM               TTL                 TTL             +12V
Signal from CRYRING ------> CAEN N89 ------> LF4000 -------> CAEN N89 -------> LeCroy 222 -------> PCB amp -------> start move in
( active high )         TTL              NIM  FIFO  -------> CAEN N89 -------> LeCroy 222 -------> PCB amp -------> start move out
                                                    not(NIM)               TTL                 TTL    ^        +12V
                                                                                                      |
                                                                                                      |
CAEN N89 mode switch to OUT                                                                                                  PSU +12V 

Widths of signals from LeCroy 222
 move in = 75ms
 move out = 25ms

Note - the outputs of EG&G/ESN LF4000 FIFO could be connected directly to the LeCroy 222 as the inputs accept NIM and TTL
Attachment 1: 20240205_174307.jpg  2.358 MB  | Hide | Hide all
20240205_174307.jpg
ELOG V3.1.4-unknown