17.43 setup signals from detector movement controller.
See attachment 1
NIM TTL TTL +12V
Signal from CRYRING ------> CAEN N89 ------> LF4000 -------> CAEN N89 -------> LeCroy 222 -------> PCB amp -------> start move in
( active high ) TTL NIM FIFO -------> CAEN N89 -------> LeCroy 222 -------> PCB amp -------> start move out
not(NIM) TTL TTL ^ +12V
|
|
CAEN N89 mode switch to OUT PSU +12V
Widths of signals from LeCroy 222
move in = 75ms
move out = 25ms
Note - the outputs of EG&G/ESN LF4000 FIFO could be connected directly to the LeCroy 222 as the inputs accept NIM and TTL |