AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CARME  ELOG logo
Message ID: 517     Entry time: Thu Feb 15 13:18:30 2024
Author: JG 
Subject: MBS DAQ: target & ring values 

We are running an MBS DAQ to record the attached list of slow control values of the ring and the target. In most cases this is achieved by converting analog monitor voltages to frequencies and recording them with a VME scaler (VULOM4B with scaler-firmware). The complete list of signals is attached including a channel mapping.

!! Target values are currently not provided, we can only record ring/cooler values !!

We do a scaler readout with a 1Hz trigger rate right now. This can be changed. This generates events with trigger number 1.

Additionally, we receive triggers from motor movement (trigger 2 & trigger 3) and from target ON and target OFF (trigger 12 & trigger 13).

Each event of any trigger type is WR timestamped.

There  is also a Go4 configured for online monitoring.

Details below. Or contact Jan Glorius.

----------------------------------------

Basic info and how-to for MBS

----------------------------------------

linux-user: cry_exp

VME CPU: r4l-58

DAQ PC: atppc025

data storage:  lxg1275:/data.local2/2024_CRYRING_16O_jet

The DAQ is running in a screen session "target_DAQ" on atppc025. Enter by typing

> screen -x "target_DAQ"

SSH connections to r4l-58 are active in tabs 0 and 1. Type [ctrl]+[a] then [x] to switch to tab x. 

If the SSH connection is lost type

> ssh cry_exp@r4l-58

> cd /esr/usr/cry_exp/2024_CRYRING_jet/mbs

MBS is running in tab 1. If it has an error, type [ctrl]+[z] to exit MBS, then

> resl

> mbs

mbs> @startup

mbs> @connect

to start a new file recording type

mbs> @runstart

If an error occures during @runstart, try @connect again.

Inside MBS you monitoring options:

mbs> show acq   (will show you trigger and data rates and status)

mbs> show file    (will show you the file status)

mbs> type ev -v 10    (will display 10 subsequent events from the data stream)

The trigger rate is currently 1Hz.

The trigger logic and setup in configured in /esr/usr/cry_exp/2024_CRYRING_jet/mbs/jet.trlo

-----------------------

Go4 monitoring

-----------------------

Go4 is running on atppc025 in the directory /u/cry_exp/Beamtimes/202402_O2+/go4_target

From this directory you can run it by typing

> go4 monitoring.hotstart

 

 

ELOG V3.1.4-unknown