Subject
*
:
14.30 per FEE64 1.8.L spectra aida11 pulser peak width 52 ch FWHM 14.45 EG&G Ortec 448 Pulser OFF all stats & histograms zero'd alpha background 22.00 DSSSD bias and leakage current OK - attachment 10 FEE64 temperatures OK - attachment 11 ADC data item stats - attachment 12 per FEE64 Stat spectra - attachment 13 Comparison of attachment 9 (per FEE64 Rate spectra with pulser) indicates significant issues (downstream of adaptor PCB) for DSSSD #1 Possible issues asics #3 aida01 and possible adaptor PCB misalignment aida05