AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
CARME |
 |
|
Message ID: 544
Entry time: Wed Apr 24 12:48:35 2024
|
Author: |
TD |
Subject: |
Wednesday 24 April |
|
|
Replace AIDA ASIC mezzanine of aida01
13.43 DSSSD bias & leakage current OK - attachment 1
ambient temperature 16.5 deg C
FEE64 temperatures OK - attachment 2
All system wide checks OK
WR timestamps OK - attachment 3
ADC data item stats OK - attachment 4
per FEE64 Rate spectra - attachment 5
c. 10 strips (of 1024) not producing data
most missing strips are from aida05 - misaligned or damaged adaptor PCB?
per p+n FEE64 1.8.L spectra - attachment 6
aida01 pulser peak width 75 ch FWHM
common x/y scale => peak height proxy for peak width
per n+n FEE64 1.8.L spectra - attachment 7
aida01 pulser peak width 70 ch FWHM
per FEE64 1.8.W spectra - 20us FSR - attachments 8-11
Summary
All FEE64s appear to be working (except aida05 which has c. 8 missing channels) with pulser - which
validates the FEE64, ASICs and adaptor PCB.
Electronic noise c. 50keV FWHM for aida01 & aida03 - other FEE64s better/worse.
Preamplifier noise appears to be similar for all FEE64s which may suggest an issue with the test signal chain
Note
1) e-target is starting up
2) have identified some instances of poor CARME cable management/routing see https://elog.ph.ed.ac.uk/CARME/545
EG&G Ortec 448 Pulser
Amplitude 50,000
Normalise 5.00
x5 attenuator IN
t_r 100ns
tau_d 50us
Polarity +
Frequency relay line (50Hz)
Cooknell SA1 Summing Amplifier
inverted output to test - daisy chain
Current ASIC settings saved to 2024Apr24-14.14.03
14.48 Pulser OFF
All histograms & stats zero'd
LEC slow comparator 0x64 -> 0xff
HEC fast comparator 0x2 -> 0x5
background alpha run to validate signal path from DSSSD to FEE64 adaptor PCB
22.18 Start Merger & TapeServer - attachments 12-14
DSSSD bias & leakage current OK - attachment 21
ambient temperature 16.0 deg C
FEE64 temperatures OK - attachment 15
WR timestamps OK - attachment 16
ADC data item stats OK - attachment 17
per FEE64 Rate spectra - attachment 18
per p+n FEE64 1.8.L spectra - attachments 19-20 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|