CARME |
 |
|
Message ID: 602
Entry time: Sat Jun 8 14:49:14 2024
|
Author: |
TD |
Subject: |
R51 - contd |
|
|
15.40 DSSSD bias & leakage current OK - attachments 1-2
log shows variations due to ambient temperature changes in CRYRING hall
FEE64 temperatures OK - attachment 3
WR timestamps OK - attachment 4
System wide checks Ok *except* global clock (aida16) - attachments 5-6
ADC data item stats OK - attachment 7
per FEE64 Rate spectra - attachment 8
per FEE64 1.8.L spectra - attachments 9-10
have not been cleared in some time probably - will zero histograms & stats for later review
per FEE64 1.8.W spectra - 20us FSR - attachments 11-12
Merger, TapeServer etc - attachments 13-14
data file G22-222/R51_15
Motion control - attachment 15
16.01 all histograms & stats zero'd
16.43 online per DSSSD energy spectra - attachment 16
photographs of CRYRING beam & target diagnostics - attachments 17-22
18.45 Left DSSSDs cycled to clear high rates (c. 250k/FEE64) aida01-aida04 per https://elog.ph.ed.ac.uk/CARME/603
18.49 DAQ continues data file G22-201/R51_27
DSSSD bias & leakage current OK - attachment 23
FEE64 temperatures OK - attachment 24
ADC data item stats OK - attachment 25
per FEE64 Rate spectra - attachment 26
online per DSSSD energy spectra - attachment 27
19.00 Beam less stable but most cycles 5e+7 or higher
Target peaking c. 1e12 dropping to c. 6e11
19.09 CRYRING hall temperature increasing, DSSSD leakage current increasing - attachments 28-29
20.06 Beam less stable most cycles 1-2e+7 or higher
Target c. 1-2e12
20.15 JM/CB advise changing Lakeshore 336 Temp Controller set point from 29k to 29.5K - done
Average pressure decreases but rapid variations still observed - see attachment 30
21.00 contacted HKR to improve beam stability - they were alreday working on it but did not know what was causing the problem - will contact on-call CRYRING expert
cycled L DSSSD to reset rates
21,10 beam 0ff - DSSDs OUT
Increased target temperature to 30K to solve instability that developed again up to E12 density range.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|