| <p>ONLY FOLLOW THESE INSTRUCTIONS if E1 turbo is below 820 Hz, pressure in E1 is above 2.5E-1 mbar and experts (Nikos, Andreas) cannot be reached.</p>
<p> </p>
<p>Go to the red cabinet under the stairs.</p>
<p>Open rightmost door. See first photo. Do not press the button on the handle or the door will lock!</p>
<p>Turn down pressure regulator indicated in second photo by approximately 1 degree. Read regulator to know which way to turn to decrease. No immediate change should be obvious on the pressure regulator.</p>
<p>Check E1 pressure for 5 minutes. If no effect, turn pressure down again until an effect appears.</p> |