AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
CARME
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Mon Mar 31 13:06:39 2025
Author
*
:
Subject
*
:
> Assembled feedthrough flange with re-cut > SS support bracket. M3 vented on the > sides, M3x30 partly threaded on top. Top > screws don't touch the > Macor (by design? See first attachment). > Macor affixed but has > some play - good for thermal dilation. > > Mounted long wire harness (side flange) on > top. Unable to get one of the two > connectors fully in. Unable to get it out > as well. See last attach, it's a small gap. > Tested few pins with multimeter - all ok. > Decided to mount. > > No issues mounting. Slowly and carefully > threaded wires harness in CARME taking > care not to hit negs. > > Slightly concerned about weight of Macor > connector on wires, but no obvious > alternative. > > Please have a look at attachments and > comment. Will keep mounting in the > afternoon.
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00