> > 12.00 CARME IE514 ion gauge 4.2e-11 mbar > CARME IE514 OFF > > > aida08 red coolant line and HDMI cable found disconnected - blue coolant line, network cable and FEE^$ power are still connected > Now re-connected > > > 13.57 DSSSD bias & leakage current - attachment 1 > leakage currents high (especially DSSSD#1 and DSSSD#3) cf. most recent values from June 2024 values https://elog.ph.ed.ac.uk/CARME/654 > another ion gauge on with line of sight to DSSSDs? > > FEE64 power ON > > FEE64 temps OK - attachment 2 > > All system wide checks OK - attachments 3-6 > > WR timestamps OK - attachment 7 > > > 18.00 DSSSD bias OFF > FEE64 power OFF