AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
CARME |
 |
|
Message ID: 668
Entry time: Tue Apr 8 10:46:13 2025
|
Author: |
TD |
Subject: |
Tuesday 8 April |
|
|
11.30 CRYRING vacuum and temperature - attachments 1-2
DSSSD bias & leakage current - attachment 3
c. 1uA reduction in DSSSD#1 and DSSSD#2 leakage currents cf. yesterday but remain high cf.June 2024
FEE64 temps OK - attachment 4
System wide checks - clock fail aida16 - attachments 5-6
aida16 global clock status 0x6 - non-critical failure, often resolved by a power cycle.
ADC data item stats - attachment 7
DSSSD #3 n+n Ohmic side FEE64 rates return to c. normal cf. yesterday
per FEE64 Rate spectra - attachment 8
per FEE64 1.8.L spectra - attachments 9-10
pulser peak width aida11 55 ch FWHM, aida09 75 ch FWHM
only DSSSD #3 aida11/aida12 have expected peak widths
per FEE64 1.8.W spectra - 20us FSR - attachments 11-13
WR timestamps OK - attachment 14
17.29 DSSSD bias & leakage current - attachment 15
FEE64 temps OK - attachment 16
WR timestamps OK - attachment 17
ADC data item stats - attachment 18
per FEE64 Rate spectra - attachment 19
17.35 All histograms & stats zero'd
Start run
Data file APR25/R5
alpha background
EG&G Ortec 448 Pulser
amplituide 90,000
normalise 5.00
x5 attenuator IN
INT OSC
frequency 2Hz
t_r 100ns
tau_d 50us (1000us setting does not work)
polarity +
ASIC settings 2025Apr08-13.45.30
slow comparator 0x64
aida03, aida04, aida09, aida10 LEC/EC range setting -> MEC
per FEE64 Rate spectra - attachment 20
Merger & Tape Server - attachment 21-22
20.41 DAQ continues
Data file APR25/R5_7
DSSSD bias & leakage current - attachment 23
FEE64 temps OK - attachment 24
ADC data item stats - attachment 25
Merger & Tape Server - attachment 26-27
per FEE64 Rate spectra - attachment 19
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|