Leak tested CARME via He leak tester + He bottle. No leaks could be found. Some indication of potential leak close to blind flange on bottom valve - disappeared after closer checks. Outgassing from cables? If so second turbo may be required for baking. Removed leak tester. Scroll on, turbo on. Pumping down overnight. Tested baking system. Missing: - Monitor (not shipped from DL??) - Monitor power cable - Power cable for USB hub (not shipped??) Could find the first two, but monitor is of poor quality. USB to RS232 cable damaged in shipping (poorly secured). Still works, may be worth repairing but not trivial. Only works plugged in to top USB 3 port on Pi. System tested and still working aside from these issues. Connected webcam to pi facing IE514 display. Will take pictures every hour on the hour as in DL. Will stop pumping tomorrow pm due to imminent shutdown. Pressure at 5E-6 mbar decreasing.