21:00
Run 35 started.
D+ beam deuterium target. D+ beam energy is 2 MeV/u. (beam calc calculation is 1.9211 MeV/u -> may be wrong cooler voltage, need to check grafana)
Target density is about 4e12 at/cm2. FLuctuations between 1e12 and 1e13 at/cm2.
All four detectors positioned fully out and stationary.
Left potentiometer = 18.86 mm
Right potentiometer = 19.8 mm
Ring cycle time is 1 minute. Note we see some hydrogen in in the spectrum, this disapears after a few seconds.
Target T1=30K, T2=31.9K.
Nozzle position x=-3.06, z=1.72 mm
YR09-Etarget pressure = 2.29e-10 mbar.
Cooler set value = 1126V, 15 mA. Read value = 1068V, 13.22 mA
Detector 4 is still a bit unstable in its leakage current but the data coming through looks good. Some crap is periodically written to disk when leakage current spikes. Attachment 1,2 for leakage currents.
Attachment 3,4 data rate to disk and statistics tab for when detector 4 is behaving. When not data rate can spike to 4000 kb/sec. Looks OK at the moment.
Attachment 5 - xy plot. We observe a circular shape on the detector. We are unsure of what this is at the moment. Offline analysis required.
Attachment 6 - energy histogram including Rutherford.
Attachment 7 - energy histogram just nuclear peaks.
Attachment 8 - beam calc with read values. |