| <p>01:10</p>
<p>Gas target turned off via the shutter valve.</p>
<p>Run 49 started beam only background run.</p>
<p>Slow comparator threshold lowered to 0x14 (200 keV) in preparation for the energy change tomorrow morning.</p>
<p>Other than that no changes to the DAQ or to the experimental setup.</p>
<p>D+ beam remains at 0.5 MeV nominally. All four detectors positioned fully out and stationary,</p>
<p>I tried to re-bias detector #4 however was not able to bias past 70V. DSSD4 remains off and set to the MEC (0x1). (see attchement 1)</p>
<p>Ring cycle is every 10 seconds. Beam position unchanged, Beam position. Vertical =0.0mm , Horizontal=0.0 mm</p>
<p>Ring vacuum is now 3.16e-11 mbar. Before switching off the gas target density was 2e12 at/cm2</p>
<p>Data rate is 89 kB/sec</p>
<p>ADC data items stats remain largely unchanged at new threshold. (see attachment 2)</p>
<p>Aida01 has increased to several thousand. Tried check ASIC control on asic #4 of aida01->no effect on higher rates.</p>
<p>01:20</p>
<p>All checks OK.</p>
<p>After 20 minutes of running see energy histogram and xy histogram (attachments 3,4)</p>
<p>Interesting xy pattern (hydrogen? Injection noise?)</p>
<p>Note put aida13-aida16 slow comparator threshold back to 0x24 as we had some rates coming through (~1kHz) which is now back to zero. If we revive the detector in the morning, the thresholds should be lowered in line with other detectors.</p>
<p>03:00</p>
<p>All checks OK</p>
<p>04:30</p>
<p>All checks ok</p>
<p>7:20</p>
<p>All checks ok</p>
<p><strong>Run stopped</strong>, final energy histogram and xy histogram attached.</p> |