11:45 New run started - run 81, same settings as previous runs (75, 76 etc)
ROOT histograms NOT zeroed
Beam at 2.5e8, target density at 3e12
11:50 All checks okay
12:45 All checks okay
13:45 All checks okay
14:45 All checks okay
15:45 All checks okay.
17:00 All Checks okay. Target temp manually dropped from 30.4K to 30.1K to try and improve density as would not get much above 2.5-3ish. Also played with target position for a while - now seemed to have stabilised it at 4.5e12 as of now.
18:00 All checkd okay. Target been stable at 4.5-5.3e12.
18:20 Done checks again (all ok) as the beam is being a little stange. Attatched ss of peaky peaks (some of them very nice) but detector 3 histogram wont cooperate and wont let me see the counts if I zoom further down, but can see proton peak fine. Seeing periodic spikes in the counts plot for the particle detector - monitoring. Also the tail of the beam seems more red than usual.
19:00 More photos of strange spikes in the counts for particle detector uploaded. Sometimes, the baseline has decent sized spikes too. Attatchments 10-14
19:20 Checks ok. Traget stable. Spikes in particle detector seem to be more frequent, uploading more photos below. Also attatched peaks at this stage (attatchemtn 15).
19:36 Very strange spikes in particle det. Two bigish peaks at baseline and then strange beginning at the injection. Attatchment 19.
19:45 Another spike in the baseline for particle det. Attatchment 20.
20:25 Checks ok. Spikes still in particle det plot.
21:20 Checks ok. Screenshot of peaks in attatchment 21.
22:20 Checks ok. Target been relativly stable all shift at 5e12. Spikes still periodically turning up in particle detector. Histograms are gaining counts, attatchment 22 is the peaks.
22:30 Target suddenly got really dense up to 1e13. Lifetime of the beam is ok. Trending down slowly. Did checks again and all is good.
01:45. Checks OK. Target seems thick.
04:30
Temperatures OK - attach
Rates OK - attach
Merger OK - attach
Hisograms look OK - attached
Target looks very thick from ring vacuum worsening.
7:15 all checks okay - historgams attached
beam 2e8, target 1e13, scintillator rate 2000 in peak
10:00
Run stopped to gain cave access.
Before beam was stopped, target was moved out. We still see some signal on the neutral particle detector, roughly a factor 10 below with target on at 1e13 at/cm2
Will need a beam background run for normalisation procedure. See attachment 29
10:30
Now we have no beam in the ring we still observe some events in the neutrals detector. Some spikes up to 150 counts observed, could explain some of what we have seen so far. See attachment 30
Note we are saving all this data to MBS so it is effectively background for the neutral detector. |