Sat Sep 12 08:38:52 2020
Author
*
:
Subject
*
:
Saturday 12 Sep 08:43 Remotely checked status of baking. See attached. Baking appears to be proceeding OK. No thermocouples disconnected. Speed is slower than previously observed, limited by thermal gradients observed across both octagon and CARME vessel as a result of new thermocouple placement. Octagon gradient may not be real, and could be a result of placement of Octagon side thermocouple? Main concern is temperature recorded by CJC in Picolog2 (aka Room temperature 2) at 31 C, increasing. The second picolog may be too close to the tent. If the temperature keeps increasing it may stop working. Will keep monitoring.