ID |
Date |
Author |
Subject |
672
|
Thu Apr 10 10:53:14 2025 |
TD | Thursday 10 April |
11.50 DAQ continues
Data file APR25/R6_13
DSSSD bias & leakage current - attachment 1
FEE64 temps OK - attachment 2
ADC data item stats - attachment 3
Merger & Tape Server - attachment 4-5
Disk file log - attachment 6
event rate increased significantly at some point this morning but is currently OK
19.21 FEE64 power OFF - aida08 high temperature
DSSSD bias & leakage current - attachment 7
FEE64 temps - attachment 8
aida08 high temperature - FEE64 power OFF, detector HV OFF |
Attachment 1: Screenshot_from_2025-04-10_11-50-32.png
|
|
Attachment 2: Screenshot_from_2025-04-10_11-51-02.png
|
|
Attachment 3: Screenshot_from_2025-04-10_11-51-35.png
|
|
Attachment 4: Screenshot_from_2025-04-10_11-51-48.png
|
|
Attachment 5: Screenshot_from_2025-04-10_11-52-02.png
|
|
Attachment 6: Screenshot_from_2025-04-10_11-53-45.png
|
|
Attachment 7: Screenshot_from_2025-04-10_19-20-23.png
|
|
Attachment 8: Screenshot_from_2025-04-10_19-20-51.png
|
|
679
|
Tue Apr 22 06:44:02 2025 |
TD | Tuesday 22 April |
07.40 DSSSD bias & leakage current - attachments 1-3
CAEN logger program crashed c. 18.00 April 20 - now restarted
DSSSD #2 and DSSSD #4 OK - DSSSD #4 stable since c. 03:00 20.4.25
DSSSD #1 and DSSSD #3 remain high with slow decrease (ambient temperature approx stable c. 16.5 deg C)
CRYRING vacuum and temperature - attachments 4-5 |
Attachment 1: Screenshot_from_2025-04-22_07-40-45.png
|
|
Attachment 2: Screenshot_from_2025-04-22_07-41-10.png
|
|
Attachment 3: Screenshot_from_2025-04-22_07-47-25.png
|
|
Attachment 4: 1.png
|
|
Attachment 5: 2.png
|
|
695
|
Tue May 27 08:38:29 2025 |
TD | [How To] Recover DAQ when one FEE64 stops sending data - short(er) version |
1) Select 'No Storage' [workspace 4, display 1, browser tab 'TapeServer']
If already in 'No Storage' mode please ignore this step
2) STOP DAQ [workspace 3, display 1, browser tab 'Run Control']
all FEE64s should change status from 'going' to 'stopped' *except* whichever FEE64(s) have a problem - typically their status is 'going' or 'error'
3) Restart Merger by selecting the Merger icon [workspace 4, display 2]
terminal window showing the startup will popup
4) Setup Merger by selecting 'SETUP' [workspace 4, display 1, browser tab 'NewMerger Control']
5) Start Merger by selecting 'GO' [workspace 4, display 1, browser tab 'NewMerger Control']
all data links will be inactive (red)
6) Start DAQ by selecting 'GO' [workspace 3, display 1, browser tab 'Run Control']
check that DAQ is *actually* going - if not repeat 'GO'
7) Check data is being transferred to Merger and TapeServer
Select 'Reload' [workspace 4, display 1, browser tab 'NewMerger Control']
you should see a non-zero 'current merged data rate' and all links should show as green at least some of the time (requires observation over several reloads)
Select 'Reload' [workspace 4, display 1, browser tab 'TapeServer']
you should see 'Blocks written' and 'kb written' counters increase
8) Unselect 'No Storage' [workspace 3, display 1, browser tab 'TapeServer']
If already in 'No Storage' mode at the start of this procedure please ignore this step
If you continue to have problems, please request expert assistance. |
708
|
Sat May 31 10:46:42 2025 |
TD | Run 38 - remote monitoring |
11.40 DAQ continues
File G22-205/R38
ASIC settings
slow comparator 0x32 all FEE64s
DSSSD bias & leakage current OK - attachments 1-2
FEE64 temps OK - attachment 3
ADC data item stats OK - attachment 4
WR timestamps OK - attachment 5
per FEE64 Rate spectra - attachment 6
Merger & Tape Server - attachments 7-8
filesystem /mnt/data1 4.7Tb remaining
Online analysis - per DSSSD energy spectra - elastic peak max count / channel c. 250-300Hz - attachment 9
CRYRING vacuum & temperature - attachments 10-12
YR09 e-target gauge indicates vacuum transients from high e-11 to low e-9mbar presumably due to high density D2 target
11.42 All histograms & stats zero'd
17.36 DSSSD bias & leakage current OK - attachments 13-14
FEE64 temps OK - attachment 15
ADC data item stats OK - attachment 16
per FEE64 Rate spectra - attachment 17
per FEE64 1.8.L spectra - attachments 18-19
aida11 p+n pulser peak width 63 ch FWHM
aida03 n+n pulser peak width 85 ch FWHM
Merger - attachment 20
Online analysis spectra - elastic peak max count / channel c. few Hz - attachment 21
rates much lower than earlier - no beam and/or no target - alert JM |
Attachment 1: Screenshot_from_2025-05-31_11-40-15.png
|
|
Attachment 2: Screenshot_from_2025-05-31_11-40-34.png
|
|
Attachment 3: Screenshot_from_2025-05-31_11-41-01.png
|
|
Attachment 4: Screenshot_from_2025-05-31_11-41-18.png
|
|
Attachment 5: Screenshot_from_2025-05-31_11-42-27.png
|
|
Attachment 6: Screenshot_from_2025-05-31_11-43-50.png
|
|
Attachment 7: Screenshot_from_2025-05-31_11-44-54.png
|
|
Attachment 8: Screenshot_from_2025-05-31_11-45-10.png
|
|
Attachment 9: Screenshot_from_2025-05-31_11-46-08.png
|
|
Attachment 10: Capture.PNG
|
|
Attachment 11: Capture_2.PNG
|
|
Attachment 12: Capture_3.PNG
|
|
Attachment 13: Screenshot_from_2025-05-31_17-26-17.png
|
|
Attachment 14: Screenshot_from_2025-05-31_17-03-49.png
|
|
Attachment 15: Screenshot_from_2025-05-31_17-03-18.png
|
|
Attachment 16: Screenshot_from_2025-05-31_17-03-38.png
|
|
Attachment 17: Screenshot_from_2025-05-31_17-26-42.png
|
|
Attachment 18: Screenshot_from_2025-05-31_17-27-39.png
|
|
Attachment 19: Screenshot_from_2025-05-31_17-27-39.png
|
|
Attachment 20: Screenshot_from_2025-05-31_17-36-49.png
|
|
Attachment 21: Screenshot_from_2025-05-31_17-37-07.png
|
|
362
|
Wed Mar 2 06:00:49 2022 |
Spillmann, Uwe | waiting |
07:00
Target density is hovering between 2.0e-8 and 2.3e-8 |
363
|
Wed Mar 2 09:38:50 2022 |
Spillmann, Uwe | ion source restarted |
detector position
out: 23.1mm
in: 67.4mm
tape: next file
gas jet S1: 2.0-2.2 ** -8
|
688
|
Fri May 23 09:19:40 2025 |
SN, TD | Target wait, switch lithium oven |
10.12 DAQ, Merger & Tape Server running OK - no storage mode
DSSSD bias leakage current OK - attachments 9-10
CRYRING vacuum & temperature - attachments 11-12
FEE64 temps OK - attachment 8
ADC & DISC data item stats OK - attachments 6-7
per FEE64 Rate spectra - attachments 1 & 5
before and after check ASIC load
WR timestamps OK - attachment 4
System wide checks OK *except* FPGA & WR status - attachments 2-3
10.33 All histograms & stats zero'd
Online analysis restarted
13:56 Checks on temperature status, statistics status and power supply : OK
|
Attachment 1: Screenshot_from_2025-05-23_10-18-44.png
|
|
Attachment 2: Screenshot_from_2025-05-23_10-15-58.png
|
|
Attachment 3: Screenshot_from_2025-05-23_10-15-54.png
|
|
Attachment 4: Screenshot_from_2025-05-23_10-15-36.png
|
|
Attachment 5: Screenshot_from_2025-05-23_10-14-27.png
|
|
Attachment 6: Screenshot_from_2025-05-23_10-13-43.png
|
|
Attachment 7: Screenshot_from_2025-05-23_10-13-32.png
|
|
Attachment 8: Screenshot_from_2025-05-23_10-13-21.png
|
|
Attachment 9: Screenshot_from_2025-05-23_10-12-52.png
|
|
Attachment 10: Screenshot_from_2025-05-23_10-12-40.png
|
|
Attachment 11: 1.png
|
|
Attachment 12: 2.png
|
|
Attachment 13: Screenshot_from_2025-05-23_12-53-29.png
|
|
Attachment 14: Screenshot_from_2025-05-23_12-53-24.png
|
|
Attachment 15: Screenshot_from_2025-05-23_12-50-59.png
|
|
Attachment 16: Screenshot_from_2025-05-23_13-53-32.png
|
|
Attachment 17: Screenshot_from_2025-05-23_13-53-16.png
|
|
Attachment 18: Screenshot_from_2025-05-23_13-52-59.png
|
|
342
|
Tue Mar 1 06:04:04 2022 |
SHB, MS | positions |
just FYI: current positions saved in files are x=1000 and y=-750. this information is good in case of crash.
|
343
|
Tue Mar 1 06:35:42 2022 |
SHB, MS | check |
detector motors are still out
'Check Clock Status', 'Check ADC Calibration' : Clock status test result: Passed 5, Failed 0
'Check the White rabbit decoder status' :
Base Current Difference
aida01 fault 0xab60 : 0xab8f : 47
aida02 fault 0x88d3 : 0x8901 : 46
aida03 fault 0x23aa : 0x23d9 : 47
aida04 fault 0x15d7 : 0x1606 : 47
aida05 fault 0x2f03 : 0x2f0f : 12
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR |
Attachment 1: Screenshot_from_2022-03-01_07-55-08.png
|
|
Attachment 2: Screenshot_from_2022-03-01_07-54-46.png
|
|
Attachment 3: Screenshot_from_2022-03-01_07-53-32.png
|
|
Attachment 4: Screenshot_from_2022-03-01_07-53-13.png
|
|
Attachment 5: Screenshot_from_2022-03-01_07-51-22.png
|
|
Attachment 6: Screenshot_from_2022-03-01_07-47-38.png
|
|
Attachment 7: Screenshot_from_2022-03-01_07-45-11.png
|
|
Attachment 8: Screenshot_from_2022-03-01_07-42-33.png
|
|
Attachment 9: Screenshot_from_2022-03-01_07-41-35.png
|
|
Attachment 10: Screenshot_from_2022-03-01_07-40-48.png
|
|
Attachment 11: Screenshot_from_2022-03-01_07-39-56.png
|
|
344
|
Tue Mar 1 08:02:26 2022 |
SHB, MS | check |
'Check Clock Status', 'Check ADC Calibration' and 'Check the White rabbit decoder status' - no changes |
Attachment 1: Screenshot_from_2022-03-01_09-08-41.png
|
|
Attachment 2: Screenshot_from_2022-03-01_09-08-28.png
|
|
Attachment 3: Screenshot_from_2022-03-01_09-08-13.png
|
|
Attachment 4: Screenshot_from_2022-03-01_09-07-55.png
|
|
Attachment 5: Screenshot_from_2022-03-01_09-07-25.png
|
|
Attachment 6: Screenshot_from_2022-03-01_09-06-12.png
|
|
Attachment 7: Screenshot_from_2022-03-01_09-05-55.png
|
|
Attachment 8: Screenshot_from_2022-03-01_09-05-42.png
|
|
Attachment 9: Screenshot_from_2022-03-01_09-04-37.png
|
|
Attachment 10: Screenshot_from_2022-03-01_09-04-04.png
|
|
Attachment 11: Screenshot_from_2022-03-01_09-03-39.png
|
|
530
|
Mon Feb 19 06:01:17 2024 |
SHB, JM, JG | Run 42 morning shift |
07:00
Observed reduction in YR09 presure over last few hours in ML online plot. Beam intensity looks OK but target density was fluctuating quite a lot. Current value was 1.5e11.
Changed the nozzle position slightly which increased target density to 5.5e11. Pressure in YR09 looks OK.
I accidently scrolled slightly with the target code (only by 100 but am not sure). Target looks stable but I will not touch remotely again for fear of losing total target position.
Rates in ADC data items statistics and in the monitoring code look consistent with beam on target.
08:56
Target density reduced again to 1.5e11. Changed the nozzle position slightly in X and recovered 5.5e11 at/cm2
Beam intensity looks OK
FEE temps OK
System checks OK
Merger and tape server OK
Statistics ADC data items and spectrum browser rates OK
Merger stats attachement 5
leakage current plot attachment 6
Online monitoring code rates OK
09:33
Run stopped to change beam energy
|
Attachment 1: Screenshot_from_2024-02-19_08-11-42.png
|
|
Attachment 2: Screenshot_from_2024-02-19_08-12-02.png
|
|
Attachment 3: Screenshot_from_2024-02-19_08-12-27.png
|
|
Attachment 4: Screenshot_from_2024-02-19_08-12-47.png
|
|
Attachment 5: Screenshot_from_2024-02-19_08-57-23.png
|
|
Attachment 6: Screenshot_from_2024-02-19_08-59-08.png
|
|
220
|
Mon Feb 21 18:03:58 2022 |
SHB, JM, CB, OH (remotely) | Info: signals on ion intensity for experiment |
Currently 4 possibilities to have a look at the beam, all of them are OK to check whether there is beam or not, but some of them are better for checking absolute values.
- Schottky RF signal
- Schottky zero span signal
- CRYRING Radio (signal from BPM) --> YR12DA1R
- ICT (integrated current transformer) --> YR12DT1C
absolute measurements:
- currently the beam is at 5.5 MeV/u calibrated on the flat top. The signal of ICT before is noisy but on the flat top it is well calibrated.
relative measurements:
- The signal from CRYRING radio, has a good amplifier and is frequency independent with a nice log signal. it can be used for relative measurements
- The zero span signal of Schottky can be used, but it is not reliable for measuring the number of ions since change in the control room may change the signal level.
- Schottky RF signal can be viewed by running Launcher-App using the button "VNC Tektronix RSA". The only problem is that only one instance can run, so if in the control room someone needs it, in the container it will disappear.
|
221
|
Mon Feb 21 20:56:51 2022 |
SHB, JM, CB, OH (remotely) | DAQ connection for ion current |
signals from ICT and CRYRADIO are now connected to AIDA2 and AIDA5 respectively. ICT is set to factor 200 and CRYRADIO to 40 for now.
(AIDA3 --> electron cooler)
|
Attachment 1: cry.jpg
|
|
222
|
Mon Feb 21 21:04:08 2022 |
SHB, JM, CB, OH (remotely) | Schottky RF view from the beam inside |
|
Attachment 1: pio-2022.02.21.21.54.46.895.PNG
|
|
580
|
Mon Jun 3 11:46:07 2024 |
SD, SHB, JM, CN, EM | Run 27 |
12:46 (attachments 1-4)
Run 27 started - Beam no target
Beam energy = 3 MeV
Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.
Beam intensity at inj ~ 2.5e7
Detectors are positioned IN pneumatic only. Left arm = 62.71, right arm = 64.60 mm
Histograms and stats zero'd, monitoring code restarted.
Beam in the ring ~ 7 minutes.
Beam on BPM looks to be better than at previous energy.
Pressure in YR09 E target ~ 4.5e-11 mbar.
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
13:45 (attachments 5-8)
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
15:15(attachments 9-12)
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
Target density 1e10 atoms/cm2. Still working on the target.
|
Attachment 1: Screenshot_from_2024-06-03_12-48-03.png
|
|
Attachment 2: Screenshot_from_2024-06-03_12-48-37.png
|
|
Attachment 3: Screenshot_from_2024-06-03_12-49-13.png
|
|
Attachment 4: Screenshot_from_2024-06-03_12-49-47.png
|
|
Attachment 5: Screenshot_from_2024-06-03_13-45-42.png
|
|
Attachment 6: Screenshot_from_2024-06-03_13-46-49.png
|
|
Attachment 7: Screenshot_from_2024-06-03_13-47-32.png
|
|
Attachment 8: Screenshot_from_2024-06-03_13-47-51.png
|
|
Attachment 9: Screenshot_from_2024-06-03_15-16-51.png
|
|
Attachment 10: Screenshot_from_2024-06-03_15-17-49.png
|
|
Attachment 11: Screenshot_from_2024-06-03_15-18-19.png
|
|
Attachment 12: Screenshot_from_2024-06-03_15-22-14.png
|
|
579
|
Mon Jun 3 08:43:22 2024 |
SD, SHB | Run 26 |
09:44
Run 26 started - Beam no target
Beam energy is being increased from 1.5 to 3 MeV
Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.
Detectors are positioned fully OUT
Histograms and stats zero'd, monitoring code restarted.
Pressure in YR09 E target ~ 4.5e-11 mbar.
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
10:48 (Attachements 1-4)
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
The detectors currently fully out for tuning.
11:57 (attachments 5-8)
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
The detectors currently fully out for tuning.
|
Attachment 1: Screenshot_from_2024-06-03_10-49-19.png
|
|
Attachment 2: Screenshot_from_2024-06-03_10-50-04.png
|
|
Attachment 3: Screenshot_from_2024-06-03_10-53-01.png
|
|
Attachment 4: Screenshot_from_2024-06-03_10-53-44.png
|
|
Attachment 5: Screenshot_from_2024-06-03_11-55-41.png
|
|
Attachment 6: Screenshot_from_2024-06-03_11-56-17.png
|
|
Attachment 7: Screenshot_from_2024-06-03_11-56-36.png
|
|
Attachment 8: Screenshot_from_2024-06-03_11-57-31.png
|
|
306
|
Sun Feb 27 14:05:30 2022 |
Rui-Jiu, Ragan | Current status |
Ring current ~7.61e7 particles/s
SW3
- Merger OK
- Screenshot uploaded
- Data rate : 1312 kbytes/s
SW4
- current plot uploaded. 3.25 uA.
SW2
- T OK
- system wide checks OK
- Check statistics tab ok
- Spectra ok
SW5
4 plots uploaded |
Attachment 1: Screenshot_from_2022-02-27_15-13-14.png
|
|
Attachment 2: Screenshot_from_2022-02-27_15-13-33.png
|
|
Attachment 3: Screenshot_from_2022-02-27_15-15-48.png
|
|
Attachment 4: Screenshot_from_2022-02-27_15-17-11.png
|
|
Attachment 5: Screenshot_from_2022-02-27_15-17-49.png
|
|
Attachment 6: Screenshot_from_2022-02-27_15-18-12.png
|
|
Attachment 7: Screenshot_from_2022-02-27_15-18-36.png
|
|
Attachment 8: Screenshot_from_2022-02-27_15-19-20.png
|
|
Attachment 9: Screenshot_from_2022-02-27_15-19-45.png
|
|
Attachment 10: Screenshot_from_2022-02-27_15-21-31.png
|
|
Attachment 11: Screenshot_from_2022-02-27_15-22-44.png
|
|
Attachment 12: Screenshot_from_2022-02-27_15-23-49.png
|
|
Attachment 13: Screenshot_from_2022-02-27_15-24-47.png
|
|
Attachment 14: Screenshot_from_2022-02-27_15-25-24.png
|
|
307
|
Sun Feb 27 15:02:16 2022 |
Rui-Jiu Chen, Yuri | Change target temperature |
16:00 change temperature from 95 k to 120k.
16:07 change temperature from 120 k to 95k.
Ring current ~4.97e7 particles/s
SW3
- Merger OK
- Screenshot uploaded
- Data rate : 1036 kbytes/s
SW4
- current plot uploaded. 3.25 uA.
SW2
- T OK
- system wide checks OK
- Check statistics tab ok
- Spectra ok
SW5
4 plots uploaded |
Attachment 1: Screenshot_from_2022-02-27_16-04-47.png
|
|
Attachment 2: Screenshot_from_2022-02-27_16-04-58.png
|
|
Attachment 3: Screenshot_from_2022-02-27_16-05-55.png
|
|
Attachment 4: Screenshot_from_2022-02-27_16-06-35.png
|
|
Attachment 5: Screenshot_from_2022-02-27_16-07-06.png
|
|
Attachment 6: Screenshot_from_2022-02-27_16-07-15.png
|
|
Attachment 7: Screenshot_from_2022-02-27_16-07-27.png
|
|
Attachment 8: Screenshot_from_2022-02-27_16-09-29.png
|
|
Attachment 9: Screenshot_from_2022-02-27_16-09-43.png
|
|
Attachment 10: Screenshot_from_2022-02-27_16-10-20.png
|
|
Attachment 11: Screenshot_from_2022-02-27_16-12-44.png
|
|
Attachment 12: Screenshot_from_2022-02-27_16-13-41.png
|
|
Attachment 13: Screenshot_from_2022-02-27_16-14-13.png
|
|
Attachment 14: Screenshot_from_2022-02-27_16-14-48.png
|
|
308
|
Sun Feb 27 15:41:33 2022 |
Rui-Jiu Chen, Yuri | current status |
16:38 change target temperature from 95 k to 140 k.
16:48 change target temperature from 140k to 95k.
Ring current ~5.62e7 particles/s
SW3
- Merger OK
- Screenshot uploaded
- Data rate : 1190 kbytes/s
SW4
- current plot uploaded. 3.25 uA.
SW2
- T OK
- system wide checks OK
- Check statistics tab ok
- Spectra ok
SW5
4 plots uploaded |
Attachment 1: Screenshot_from_2022-02-27_17-01-29.png
|
|
Attachment 2: Screenshot_from_2022-02-27_17-02-11.png
|
|
Attachment 3: Screenshot_from_2022-02-27_17-04-47.png
|
|
Attachment 4: Screenshot_from_2022-02-27_17-05-14.png
|
|
Attachment 5: Screenshot_from_2022-02-27_17-05-33.png
|
|
Attachment 6: Screenshot_from_2022-02-27_17-05-44.png
|
|
Attachment 7: Screenshot_from_2022-02-27_17-06-47.png
|
|
Attachment 8: Screenshot_from_2022-02-27_17-07-02.png
|
|
Attachment 9: Screenshot_from_2022-02-27_17-07-48.png
|
|
Attachment 10: Screenshot_from_2022-02-27_17-08-29.png
|
|
Attachment 11: Screenshot_from_2022-02-27_17-09-03.png
|
|
Attachment 12: Screenshot_from_2022-02-27_17-09-36.png
|
|
Attachment 13: Screenshot_from_2022-02-27_17-10-26.png
|
|
309
|
Sun Feb 27 16:30:35 2022 |
Rui-Jiu Chen, Yuri | Change target temperature again |
1. change T from 95 k to 140 k.
2. keep T at 140k for 15 minutes.
3. change T from 140 k to 95 k. |