Wed Apr 9 10:02:21 2025, TD, CARME equipment in CRYRING area moved 
|
There was some tidying up of the CRYRING hall in advance of this week's EC evaluation meeting at GSI.
|
Thu Apr 10 10:53:14 2025, TD, Thursday 10 April 8x
|
11.50 DAQ continues
Data file APR25/R6_13
|
Tue Apr 22 06:44:02 2025, TD, Tuesday 22 April    
|
07.40 DSSSD bias & leakage current - attachments 1-3
CAEN logger program crashed c. 18.00 April 20 - now restarted
|
Tue May 27 08:38:29 2025, TD, [How To] Recover DAQ when one FEE64 stops sending data - short(er) version
|
1) Select 'No Storage' [workspace 4, display 1, browser tab 'TapeServer']
If already in 'No Storage' mode please ignore this step
|
Sat May 31 10:46:42 2025, TD, Run 38 - remote monitoring 21x
|
11.40 DAQ continues
File G22-205/R38
|
Wed Mar 2 06:00:49 2022, Spillmann, Uwe, waiting
|
07:00
Target density is hovering between 2.0e-8 and 2.3e-8 |
Wed Mar 2 09:38:50 2022, Spillmann, Uwe, ion source restarted
|
detector position
out: 23.1mm |
Fri May 23 09:19:40 2025, SN, TD, Target wait, switch lithium oven 18x
|
10.12 DAQ, Merger & Tape Server running OK - no storage mode
|
Tue Mar 1 06:04:04 2022, SHB, MS, positions
|
just FYI: current positions saved in files are x=1000 and y=-750. this information is good in case of crash.
|
Tue Mar 1 06:35:42 2022, SHB, MS, check 11x
|
detector motors are still out
'Check Clock Status', 'Check ADC Calibration' : Clock status test result: Passed 5, Failed 0
'Check the White rabbit decoder status' : |
Tue Mar 1 08:02:26 2022, SHB, MS, check 11x
|
'Check Clock Status', 'Check ADC Calibration' and 'Check the White rabbit decoder status' - no changes |
Mon Feb 19 06:01:17 2024, SHB, JM, JG, Run 42 morning shift 6x
|
07:00
Observed reduction in YR09 presure over last few hours in ML online plot. Beam intensity looks OK but target density was fluctuating quite a
lot. Current value was 1.5e11. |
Mon Feb 21 18:03:58 2022, SHB, JM, CB, OH (remotely), Info: signals on ion intensity for experiment
|
Currently 4 possibilities to have a look at the beam, all of them are OK to check whether there is beam or not, but some of them are better for checking
absolute values.
- Schottky RF signal |
Mon Feb 21 20:56:51 2022, SHB, JM, CB, OH (remotely), DAQ connection for ion current
|
signals from ICT and CRYRADIO are now connected to AIDA2 and AIDA5 respectively. ICT is set to factor 200 and CRYRADIO to 40 for now.
(AIDA3 --> electron cooler) |
Mon Feb 21 21:04:08 2022, SHB, JM, CB, OH (remotely), Schottky RF view from the beam inside
|
|
Mon Jun 3 11:46:07 2024, SD, SHB, JM, CN, EM, Run 27 12x
|
12:46 (attachments 1-4)
Run 27 started - Beam no target
Beam energy = 3 MeV |
Mon Jun 3 08:43:22 2024, SD, SHB, Run 26 8x
|
09:44
Run 26 started - Beam no target
Beam energy is being increased from 1.5 to 3 MeV |
Sun Feb 27 14:05:30 2022, Rui-Jiu, Ragan, Current status 14x
|
Ring current ~7.61e7 particles/s
SW3
- Merger OK |
Sun Feb 27 15:02:16 2022, Rui-Jiu Chen, Yuri, Change target temperature 14x
|
16:00 change temperature from 95 k to 120k.
16:07 change temperature from 120 k to 95k.
|
Sun Feb 27 15:41:33 2022, Rui-Jiu Chen, Yuri, current status 13x
|
16:38 change target temperature from 95 k to 140 k.
16:48 change target temperature from 140k to 95k.
|